The AS-1 Seismograph – Installation
and Calibration 1 L. Braile,
November, 2002; Updated
November 2004 http://web.ics.purdue.edu/~braile |
|
|
Introduction:
The AS-1 Seismometer (http://www.amateurseismologist.com/,
http://jclahr.com/science/psn/as1/) and the data acquisition software
AmaSeis (http://www.geol.binghamton.edu/faculty/jones/), written by Dr. Alan Jones, SUNY,
[1] Last
modified
The web page for
this document is:
http://web.ics.purdue.edu/~braile/edumod/as1mag/as1mag1.htm.
Partial support for this development provided by IRIS and the National Science Foundation.
ã Copyright 2002-4. L. Braile. Permission granted for reproduction for non-commercial uses.
Figure 1. The AS-1 seismometer showing the main
components of the instrument. The AS-1
is a vertical component seismometer. Up
and down motions of the ground, and therefore of the base and frame of the
seismometer, cause the coil to move relative to the magnet that is suspended by
the spring and boom assembly. The mass
of the seismometer, consisting primarily of the magnet, boom and the washers,
tends to remain steady because of inertia when the base moves. The motion of the coil relative to the magnet
generates a small current in the coil.
The current is amplified and digitized by an amplifier unit (not shown)
and connected to the computer for recording and display. The damping (using oil in the container and a
washer mounted to a bolt extending downward from the boom into the oil) reduces
the tendency for the mass and spring system to oscillate for long duration from
a single source of ground motion (arrival of seismic waves at the location).
Example Seismograms:
When a very large
earthquake occurs anywhere in the world (or a smaller event that is located
closer to the seismograph station), the AS-1 seismometer responds to the small
ground vibrations that are generated by the earthquake and propagate from the
earthquake location to the seismograph station.
The ground vibration is converted by the seismometer to a small
electrical current that is converted to a digital signal that is sent to an
attached computer and the AmaSeis software.
An example of a recorded earthquake is shown in Figure 2. This record is from a very large event
located about 42 degrees (4670 km) away from the AS-1 seismograph in West
Lafayette, Indiana.
Figure 2. Screen image of 24-hour AS-1/AmaSeis record
for
A close-up view of the
Figure 3. Close-up of
Although the AS-1 seismograph is
a relatively simple and inexpensive seismograph, it is capable of relatively
accurate and effective recording of ground vibration from earthquakes and other
sources. A comparison between an AS-1
seismogram for the
Figure 4. Comparison of the first 12 minutes of the
November 3, 2002 central Alaska earthquake seismogram (upper trace; band pass
filtered from 0.001 – 0.1 Hz in AmaSeis) recorded by an AS-1 seismograph at
West Lafayette, Indiana with a similar seismogram (lower trace; band pass
filtered from 0.1 – 2 Hz in AmaSeis) recorded by a Guralp high-performance, broadband,
digital seismograph at West Lafayette, Indiana.
Both seismograms are vertical component records. The seismograph stations are about 5 km
apart.
To see an online AS-1 record from
http://www2.bc.edu/~kafka/Seismograph_Display/EarthMoves.html. For more information on this online seismograph display, see
http://www2.bc.edu/~kafka/AS1_Web_Tool/Seismo_Display.html. This site also includes links to some sample recordings of earthquakes at the Boston College AS-1 seismograph stations. The WLIN near-real time seismograph display can be viewed at:
http://web.ics.purdue.edu/~braile/AS1Display/AS1display.jpg.
Notes on Installing the AS-1 Seismograph and the
AmaSeis Software:
Installing the AS-1 Seismometer:
Local earthquakes (0 – 5 degrees distance) – magnitude ~3.5+
Regional earthquakes (5 – 20 degrees distance) – magnitude ~4.5+
Distant earthquakes (20 – 90 degrees distance) – magnitude ~6.0+
Very distant earthquakes (90 – 180 degrees distance) – magnitude ~6.5+
(these “rules of thumb” will depend on the site and gain level, your location, for example, western US or eastern US, and the nature of the earthquake and the path that the waves have traveled from the earthquake location to your station).
My experience is that a very “visible” earthquake signal is recorded, on the average, of about once a week. About half of these can be used to calculate a magnitude (mb or MS) and about a fourth are at distances of less than 105 degrees and have S arrivals that are distinct enough (usually after filtering with a 0.001 to 0.2 Hz bandpass filter in AmaSeis; the S arrival is often best seen by the lower frequency arrival) to pick an S-P time (using the arrival picking tool in AmaSeis) and estimate the distance to the event using the S-P and travel time curve matching tool in AmaSeis.
Setting up the AmaSeis software:
1. Help file for AmaSeis:
AmaSeis (http://www.geol.binghamton.edu/faculty/jones/), has a fairly complete help file that includes instructions for setting up the software, options that are available for analyzing and displaying data, and a tutorial. To obtain help information, click on the Help icon (menu) at the top of the screen, and then on Help topics. Select Contents. Double-click on one of the topics and then on one of the question marks that appears before the name of a sub-topic, then use the double-arrow controls (right or left arrows at the top of the Help dialog box) to find information related to that topic. To investigate a specific topic, feature or setting, select Index from the Help topics dialog box, and double-click on the appropriate entry. An online tutorial with detailed descriptions for the use of the AmaSeis software is also available at:
http://web.ics.purdue.edu/~braile/edumod/as1lessons/UsingAmaSeis/UsingAmaSeis.htm.
2. Settings For AmaSeis:
Settings Menu
Helicorder
Lines per hour: usually set to 1 (gives a 24 hour record on screen which is useful for monitoring because when you observe the screen once per day, you will know immediately if an earthquake has occurred; to go back to look at earlier data, use the scroll bar on the right); can be set to as high as 20 (60 in the latest release of AmaSeis) so that the cursor moves quickly across the screen and the details of the waveform can be seen; this setting is useful for laboratory testing and demonstrations and “make your own earthquake” (stomp test, etc.) activities; for lines per hour greater than 1, temporarily set the decimate factor to 1.
Gain: amplification of the signal by the software; set to 1 for set up and initial centering of the trace on the screen, then to a higher number depending on the noise level; in reasonably quiet installations, a gain of 50 or more is possible.
Low pass filter cutoff: low pass filter option for the trace on the helicorder screen display; set at 3 Hz; if high frequency noise is visible on the screen, try a lower cutoff such as 1 or 0.5Hz to see if the noise is reduced; earthquake signals will generally not be greatly affected by this cutoff frequency for this seismometer.
Days to retain record: this value controls how far back in time one can scroll (if you are scrolling back more than a few days, move the scroll bar slider with the mouse instead of using the up or down arrows or clicking on the bar above or below the slider) to view the seismic data on the screen; set to 365 to allow viewing data back as far as one year; the value can be set larger, but if there are old data that you would like to view, they usually would have been saved as individual events (extracted and saved as .sac files; so that data archive doesn’t get too large) and then opened under the File menu.
Decimate factor: Set to 10, normally; this setting will provide a fairly accurate view of the data on the screen but allow rapid scrolling; a decimate factor of 10 plots only every 10th point; some distortion is caused by this choice, so for large numbers of lines per hour, “make your own earthquake” experiments, and whenever you want to see the most accurate view of a seismogram on the screen, set the decimate factor to 1. The decimate factor does not affect the plotting of an extracted seismogram which is always plotted with all of the available points.
Settings
Show data values
This dialog box and display allows one to center the cursor so that it tracks approximately on the “zero line” (blue line) that is defined for each hour. To center the trace, it is convenient to start with a low gain setting (such as 1). Turn the black knob on the small black interface box that comes with the AS-1 seismometer that between the seismometer and the computer in the electric circuit. The knob adjusts the cursor position on the screen. It is very sensitive, so make small adjustments and determine which way to turn the knob in order to cause the cursor to move upwards. As you adjust the knob, the cursor should get closer to the zero line and the display numbers should be close to zero. When you have succeeded in getting the numbers close to zero, turn the gain up and repeat, eventually setting the gain to the highest level that is consistent with the background noise (you should see relatively continuous, small, about +/- 1 to 2 mm, movements on the screen), and making the displayed values close to zero. To adjust the knob when the gain is set to a high value (greater than about 10), it is convenient to use a pencil and very, very lightly tap the knob from the side to cause it to move. THIS CENTERING PROCESS CAN TAKE CONSIDERABLE TIME – PROCEED SYSTEMATICALLY AND PATIENTLY.
Recording the AS-1 seismic data with the AmaSeis software generates about 1 MB of data per day. If you computer has insufficient disk space to archive data for many months or years, you can copy the files onto a zip disk or to a CDR. Alternatively, you can sort through the recorded data and delete data for days in which there was no significant signal. An example of a 24-hour record from the AS-1 seismometer and the AmaSeis software is shown in Figure 5.
Figure 5. Example of screen image. Gain setting at 60 (relatively quiet site in
basement, on concrete slab). Note noise
level. The event at about 7 hours and 10
minutes is a magnitude 6.7 earthquake near the north
4. Time Synchronization and Maintaining a
Seismograph Catalog:
It
is useful and “good scientific practice” to develop a catalog for your
seismograph station. Entries are made to
the catalog every day or two that include absolute time checks or
synchronizations (comparing the computer clock display to an accurate, absolute
time signal such as from a radio-synchronized clock (arctime.com) or radio station WWV, 5, 10 or 15
MegaHertz frequency, on a short wave radio; alternatively, if the computer is
connected to the internet, the computer clock can be periodically synchronized
to absolute time using the software tool AboutTime, http://www.arachnoid.com/), times of
arrivals (and amplitude and magnitude information) of earthquakes that are
recorded by the seismograph, and comments about the weather or background noise
conditions or operational status of the seismograph.
A
very convenient radio synchronized clock (“Atomic Clock”) is the SkyScan wall
or desk clock (I purchased one at our Sam’s Club store for $20; Figure 6) that
has very large lettering. The clock can
be set to show GMT time and can be used to visually check the time correction
of your computer running AmaSeis and to record the “drift” of the computer
clock. Occasionally resetting
(“synchronizing”) the computer clock to GMT time will keep the AmaSeis time
display reasonably accurate. A time
correction can be applied to saved seismogram files in AmaSeis. The atomic clock makes a nice display for GMT
time even if you have an Internet connection on your computer that is used to
run AmaSeis and can update your computer clock automatically using the Internet
tools described above. The SkyScan
Atomic Clock is also available ($30) at: http://eggshop.net/skysatcloc.html, or, http://homestore3.com/skysatcloc.html.
Another atomic clock option (with large digits display) is the WS-8001U
wall clock from La Crosse Technology, available at: http://www.weathermeter.com/ ($40).
Figure 6. SkyScan atomic
clock. A label describing GMT or UTC
time was added.
Relative Calibration of
the AS-1 Seismograph: A relative calibration of the AS-1
seismograph can be conducted using a “step function” method. A step function (a small, sudden increase or
decrease in the position of the mass with respect to the coil) can be applied
to the seismometer by lowering (or removing) a small mass onto the boom of the
instrument. Because lowering the mass
causes an inconsistent input to the seismometer, the method works best by
lifting the mass. Because the
seismometer is very sensitive, the relative calibration process, although simple,
must be performed with a very specific procedure and very carefully. To perform the calibration, use a piece of
masking tape or post-it note material to mark a position on the boom of the
AS-1 seismometer that is 10 cm from the hinge and upright. Place the cover back over the instrument and
drill a small hole (~4mm or 1/8 in) in the top of the cover above the mark on
the boom. Make the mass used for
calibration by cutting a 1 x 2 cm rectangle of lightweight poster board. This mass is about 0.063 g. Fold the poster board rectangle into the
shape of a “tent” and attach a 3 m long piece of nylon monofilament thread
(very light; available in fabric and sewing stores) to the top of the poster
board tent with a small piece of tape.
Thread the monofilament through the hole in the cover (from the inside
of the cover) so that the small mass is suspended above the mark on the boom by
the thread. Because the instrument is
sensitive to tilt if one is standing near it, and because the air current noise
is large, put the cover on, and lower the mass through the small hole in the
top of the cover by holding the other end of the thread (it is convenient to
attach a piece of tape to the free end of the thread to make it easier to find
and hold on to the end of the thread) from a distance of about 3 m away,
standing still. Lowering the mass gives
a calibration pulse, but a better pulse (about 1400-1500 counts maximum
amplitude) is obtained by letting the instrument stabilize and then pulling on
the thread to lift the mass suddenly off the boom. This method appears to be stable and results
in calibration pulses (Figure 7) that varied only about +/- 10% between
tests. The maximum amplitude of this
test should provide a relatively linear correction factor to the amplification
of different instruments. For more
information on calibration of the AS-1 seismometer, see http://quake.eas.gatech.edu/MagWeb/CalReptAS-1.htm.
This relative
method of calibration has the following uses:
1. Performing the calibration tests to see if
the seismometer is working properly (the calibration pulse should look
approximately like the one in Figure 7; the “lifting the mass” pulse is used;
the calibration pulse is extracted from the standard AmaSeis display to view it
in a close-up view). Repeated
calibrations suggest that measurable values of the calibration test pulse
should be (approximately):
amplitude of first trough
(negative) = -1420 +/- 150 counts
amplitude of first peak
(positive) = 480 +/- 50 counts
time from beginning of calibration pulse to
first zero crossing = 3.0 +/- 0.5 seconds
2. The relative calibration allows one to
compare different instruments. AS-1
seismographs that have calibration pulses that are approximately the same as
the pulse in Figure 7 can be assumed to have the same displacement
amplification values as given here, and can be used to estimate magnitudes
using the amplification values and procedures described here. If the calibration pulse for a different AS-1
seismometer is similar to the one shown in Figure 7 (and parameters given
above) but has different maximum amplitude, one can adjust the amplification
factors used in the magnitude calculations by the relative difference between
the two AS-1 instruments.
3. The calibration provides a check on the
polarity of the seismometer. Because
lifting the mass from the boom causes the boom to move up relative to the coil
that is attached to the base of the seismometer, the first motion of the output
trace (the calibration pulse) should be down or negative. (A first “up” motion
of the ground from an incoming seismic wave will move the base and the coil up
relative to the magnet [corresponding to a relative down motion of the boom]
attached to the boom, because of inertia of the seismometer mass suspended by
the spring. This up motion of the ground
will therefore cause an up or positive signal on the seismograph trace.) If the calibration pulse (corresponding to
lifting the small mass) has a first positive motion, the polarity of the
seismometer is reversed. To correct the
polarity, simply switch the input wire connections on the AS-1 amplifier box.
Figure 7. Calibration pulse created by suddenly lifting
a small mass (0.063 g) from the boom (at a location 10 cm from the hinge) of
the AS-1 seismometer. Enlarging (by
selecting and extracting in the AmaSeis program) the calibration pulse allows
one to measure the characteristic parameters of the pulse (maximum amplitude of
the two peaks of the pulse and time between the start of the pulse and the
first zero crossing) for comparison between instruments.
Additional information on installation and operation of an AS-1 seismograph is available at: http://www.scieds.com/spinet/AS1AmaSeis_v2.pdf (a complete manual for setting up the seismograph, using the AmaSeis software and interpreting the recorded seismic data).
Return to
Braile’s Earth Science Education Activities page
Related
Pages:
The AS-1
Seismograph – Operation, Filtering, S-P Distance Calculation, and Ideas for
Classroom Use
The AS-1
Seismograph –Magnitude Determination